Our Products

Uncalibrated Intrasense

The IntraSense series absolute pressure sensors are designed to fit into a 1-French hypo tube. The sensor comes pre-attached to cabling, simplifying the connection for the end user. The fully encapsulated electronics allow the device to be used without additional gel or encapsulant. This sensor compares pressure in vivo to an onboard vacuum cavity for reference to an absolute standard. It delivers accurate and stable pressure for acute procedures in the clinically useful range of -300mmHg to +500mmHg (460mmHg to 1260mmHg absolute) and from 10°C to 60°C. The output is stable in 37°C saline, with typical drifts of <2mmHg absolute maximum per 24 hours. The device is available either as a sensor plus cable, or with an optional 4-pin PCB board attached to the proximal end.

Technical Data

Key Specifications

  • Miniature sized sensor: 750μm × 220μm × 75μm
  • Pressure type: absolute
  • Pressure range: 460mmHg to 1260mmHg
  • Connected to 3-strand insulated cable
  • Fits within 1-French catheter tubes
  • Fully encapsulated for ease of integration
  • Millivolt output
  • Compliant with applicable AAMI/ANSI BP22 performance standards
  • For acute procedures (up to 24 hours)
  • RoHS and REACH Compliant
  • Biocompatible materials
  • Optional Lightshield

Product Configurations

Product Description Part Number Light Sensitivity Pressure Sensitivity Wire Length
IntraSense 3-wires 60cm SMI-1A-48-060-BAUU 25 mmHG 5 uV/V/mmHg 60 cm
IntraSense 4-pin PCB 60cm SMI-1A-48-060-BBUU 25 mmHG 5 uV/V/mmHg 60 cm
IntraSense 3-wires 100cm SMI-1A-48-100-BAUU 25 mmHG 5.5 uV/V/mmHg 100 cm
IntraSense 4-pin PCB 100cm SMI-1A-48-100-BBUU 25 mmHG 5.5 uV/V/mmHg 100 cm
IntraSense 3-wires 180cm SMI-1A-48-180-BAUU 25 mmHG 5 uV/V/mmHg 180 cm
IntraSense 4-pin PCB 180cm SMI-1A-48-180-BBUU 25 mmHG 5 uV/V/mmHg 180 cm
IntraSense 3-wires 60cm, light shield SMI-1B-48-060-BAUU 10 mmHG 5.5 uV/V/mmHg 60 cm
IntraSense 4-pin PCB 60cm, light shield SMI-1B-48-060-BBUU 10 mmHG 5.5 uV/V/mmHg 60 cm
IntraSense 3-wires 100cm light shield SMI-1B-48-100-BAUU 10 mmHG 5 uV/V/mmHg 100 cm
IntraSense 4-pin PCB 100cm light shield SMI-1B-48-100-BBUU 10 mmHG 5 uV/V/mmHg 100 cm
IntraSense 3-wires 180cm, light shield SMI-1B-48-180-BAUU 10 mmHG 5.5 uV/V/mmHg 180 cm
IntraSense 4-pin PCB 180cm, light shield SMI-1B-48-180-BBUU 10 mmHG 5.5 uV/V/mmHg 180 cm

Applications

  • Embolization
  • Hemodynamics / Thermodilution
  • Intracranial Pressure
  • Compartment Syndrome
  • Atrial Ablation
  • Microvascular Obstruction
  • Fractional Flow Reserve (FFR)
  • Endourology
  • Atherectomy
  • Animal testing
  • Reproductive Health
  • Glaucoma
  • Hemorrhage control
  • Endoscopy
  • Aortic Control
  • Cochlear Implant

Downloads

Uncalibrated Intrasense datasheet (pdf)

Intrasense: Uncalibrated Vs. calibrated series (pdf)

Safety considerations for medically invasive sensors (pdf)

Similar Articles

Product Inquiry

Please contact us. We will get back to you as soon as possible.
Product Inquiry

Product
Uncalibrated Intrasense

I have taken note of the privacy policy. Note: You can revoke your consent at any time for the future by sending an email to info@mev-elektronik.com.

Your Contact Person

Pascal Wittfeld

Field Application Engineer

+49 (0)5424 2340-54

write email
Contact us!
Please enter one ore more search terms.
crossmenu linkedin facebook pinterest youtube rss twitter instagram facebook-blank rss-blank linkedin-blank pinterest youtube twitter instagram