Our Products

Pressure sensors in bypass configuration for gas flowmeters

Mass flow and volumetric flow Q
A restrictive element in the main channel defines the relationship between gas flow F and differential pressure (∆P):
F = ƒ (∆P)

Typically, the gas flow F is measured as mass flow [mass per time]. If needed, volumetric flow Q [volume per time] can be derived from mass flow.
The volumetric flow is equal to the mass flow over gas density:

Q =
/ ρ;

From the Ideal Gas Law, the gas density can be found as:
ρ
= (MP) / (RT)

 

Standard volumetric flow Qs
Standard volumetric flow is a volumetric flow defined at “standard” tem
perature (T
std) and “standard” pressure (Pstd). Different manufacturers
refer to different standards (e.g., T
std = 21.1 °C or 70 °F, Pstd = 101.3 kPa or
14.7 psia).

Commonly used units for standard volumetric flow are “standard liters per
minute [slm]” or “standard cubic centimetres per minute [SCCM]”.

For a given gas, volumetric flow at non-standard temperature (T) and
non-standard pressure (P) can be found as:

Q = Qs (Ps/P) (T/Ts)

Laminar and orifice-like flow restrictive elements

Ideally, a pressure drop on a laminar restrictive element increases
linearly with the flow, while a pressure drop on an orifice increases

quadraticly (Figure 2).

While the production cost of a laminar restrictive element is higher, it

has two advantages in comparison to an orifice-like restrictor:

wider flow measurement range (F2 > F1);
increased sensitivity around zero flow.
In reality, a flow restrictive element is a combination of the two restric

tors described above; either the linear or quadratic pressure-from-flow

characteristic dominates.

Definitions:
P: pressure drop on a flow-restrictive element;
: mass flow;
Q: volumetric flow;

ρ
: gas density;
M: molar mass;

P: pressure;

R: gas constant;

T: absolute temperature

Barometric correction

For any thermo-anemometer type differential pressure sensor, including
the LDE/LME/LMI, output signal V
out is proportional to gas density ρ.
That is why barometric correction is required for
P measurements.

Vout ~ P · ρ (1)

From Poiseuille’s equation, pressure drop on a laminar restrictor P is
proportional to mass flow
and inversely proportional to gas density ρ:

P ~ [μL/D4] · · 1/ρ (2)

From (1) and (2)
V
out ~ [μL/D4] · (3)

From Bernoulli’s equation pressure drop on an orifice-like restrictor P
is proportional to mass flow in power of two
2 and inversely proportio
nal to gas density
ρ:

P ~ [1/D4] · 2 · 1/ρ (4)
From (1) and (4)

V
out ~ [1/D4] · 2 (5)

From (3) and (5) follows that the LDE/LME/LMI sensors intrinsically
require no barometric correction for mass flow measurements.

Definitions:
P: pressure drop on a flow-restrictive element;
: mass flow;
ρ
: gas density;
μ
: gas viscosity;
L: length of a flow-restrictive element;

D: inner diameter of a flow-restrictive element

Temperature compensation

The LDE/LME/LMI families feature an embedded temperature sensor.
Depending on the application, the LDE/LME/LMI sensor can be fully tem

perature compensated at the factory either for mass flow or for differential

pressure.

Bypass flow

A main channel restrictor’s pressure/flow characteristic is usually
defined without considering bypass flow and bypass flow variation from

sample to sample. Thus, a smaller flow in the bypass results in better

bypass/main channel split ratio and therefore higher accuracy. The

amount of flow in the bypass channel is defined by a sensor’s pneumatic

impedance Zp [pressure per flow]. The higher the impedance, the lower

the bypass flow. The pneumatic impedance of the LDE/LME/LMI families

can be found in a range from 10,000s to 100,000s (Pa · s) / (ml).

For example, if a 250 Pa LDE sensor’s pneumatic impedance Zp is

25,000 (Pa · s) / (ml), then the bypass flow at nominal pressure F
250 can
be found as

F
250 = P / Zp = 250 Pa / 25,000 (Pa · s) / (ml) = 0.01 mL/s.

LDE/LME/LMI features suitable for flow metering

No temperature or barometric compensation is needed for mass flow
application.

The highest-in-class pneumatic impedance guarantees the highest im-
munity to contamination and the highest bypass/main channel split ratio.

An embedded temperature sensor can be read out by the user for
temperature correction in volumetric flow application (see paragraph 1.2).

Linearized sensor output is convenient for expanding pressure and
flow dynamic range by “cascading” the LDE/LME/LMI sensors. For

example, a 50 Pa sensor can be read out in parallel with a 500 Pa

sensor virtually without data irregularities when transitioning from

sensor to sensor.

Technical Data

  • No temperature or barometric compensation is needed for mass flow application.
  • The highest-in-class pneumatic impedance guarantees the highest immunity to contamination and the highest bypass/main channel split ratio.
  • An embedded temperature sensor can be read out by the user for temperature correction in volumetric flow application.
  • Linearized sensor output is convenient for expanding pressure and flow dynamic range by “cascading” the LDE/LME/LMI sensors. For example, a 50 Pa sensor can be read out in parallel with a 500 Pa sensor virtually without data irregularities when transitioning from sensor to sensor.

Product Inquiry

Please contact us. We will get back to you as soon as possible.
Product Inquiry

Product
Pressure sensors in bypass configuration for gas flowmeters

I have taken note of the privacy policy. Note: You can revoke your consent at any time for the future by sending an email to info@mev-elektronik.com.

Your Contact Person

Pascal Wittfeld

Field Application Engineer

+49 (0)5424 2340-54

write email
Contact us!
Please enter one ore more search terms.
crossmenu linkedin facebook pinterest youtube rss twitter instagram facebook-blank rss-blank linkedin-blank pinterest youtube twitter instagram